RP-HPLC Method Development and Validation of Bicalutamide and Atenolol in its Pure Dosage Forms Bichala Prem Kumar1,*, Suthakaran R.2, Shankar Ch.3, Ghouse Syed3, Beula Jannet3, Naresh R.3 1Associate Professor, Department of Pharmaceutical Analysis, Vijaya College of Pharmacy, Munaganoor-501511, Hyderabad, Telangana, India 2Professor and Principal, Department of Pharmaceutical Chemistry, Vijaya College of Pharmacy, Munaganoor-501511, Hyderabad, Telangana, India 3Associate Professor, Department of Pharmaceutical Analysis, Vijaya College of Pharmacy, Munaganoor-501511, Hyderabad, Telangana, India *Corresponding Author E-mail: prembichala@gmail.com
Online published on 3 June, 2020. Abstract A new method was established for simultaneous estimation of Bicalutamide and Atenolol by RP-HPLC method. The chromatographic conditions were successfully developed for the separation of Bicalutamide and Atenolol by using Agilent C18 5μm (4.6*250mm) column, flow rate was 1ml/min, mobile phase ratio was Water: ACN (70: 30%v/v), detection wave length was 254nm. Top Keywords Agilent C18, Bicalutamide and Atenolol, RP-HPLC method. Top |