(54.237.183.249)
Users online: 2721    [ij] [ij] [ij] 
Email id
 

Asian Journal of Pharmacy and Technology
Year : 2017, Volume : 7, Issue : 3
First page : ( 153) Last page : ( 156)
Print ISSN : 2231-5705. Online ISSN : 2231-5713.
Article DOI : 10.5958/2231-5713.2017.00025.3

RP-HPLC Method Development and Validation for Estimation of Sofosbuvir in Pure and Tablet Dosage Form

Swathi P.*, Dutt K. Rajeswar, Rao K. N. V, Raja M. Alagar

Nalanda College of Pharmacy, Charlapally, Nalgonda, Telangana.

*Corresponding Author E-mail:

Online published on 17 October, 2017.

Abstract

A rapid and precise Reverse Phase High Performance Liquid Chromatographic method has been developed for the validation of Sofosbuvir, in its pure form as well as in tablet dosage form. Chromatography was carried out on a Hypersil C18 (4.6×150mm, 5μ) column using a mixture of Methanol (100% v/v) as the mobile phase at a flow rate of 1.0 mL/min, the detection was carried out at 265 nm. The retention time of the Sofosbuvir was 3.515 ±0.02min. The method produce linear responses in the concentration range of 20–100μg/mL of Sofosbuvir. The method precision for the determination of assay was below 2.0% RSD. The method is useful in the quality control of bulk and pharmaceutical formulations.

Top

Keywords

Sofosbuvir, RP-HPLC, Validation, Tablet dosage forms.