(18.207.102.38)
Users online: 1738    [ij] [ij] [ij] 
Email id
 

Asian Journal of Pharmaceutical Analysis
Year : 2017, Volume : 7, Issue : 2
First page : ( 87) Last page : ( 92)
Print ISSN : 2231-5667. Online ISSN : 2231-5675.
Article DOI : 10.5958/2231-5675.2017.00015.1

Method Development and Validation of Valsartan and Its Impurities by High Performance Liquid Chromatography

Chandana O. S. S.*, Babu R. Ravichandra

Department of Chemistry, Institute of Science, GITAM University, Visakhapatnam, Andhra Pradesh, India 530045

*Corresponding Author E-mail: osschandana@gmail.com

Online published on 18 September, 2017.

Abstract

Our main objective is to develop an accurate and precise HPLC method for the determination of valsartan and its impurities. An inertsil ODS-3v; (150 × 4.6) mm; 5 μm column is used for the Separation of drugs. The flow rate was maintained at 1.0 mL/min and the wave length used for detection was 230 nm. The linearity was observed in the range of 0.025–50μg/ml of spiked impurities in valsartan, impurity B and impurity C with a correlation coefficient of 0.990, 0.996 and 0.997 respectively. Linearity, accuracy, precision and robustness parameters for the suggested method were estimated for validation. The developed method can be utilized in the analysis of valsartan tablets.

Top

Keywords

HPLC, Method validation, related impurities, entacapone, LOQ, LOD.